Instrument Function
R&S
®
SMJ100A
283
Operating Manual 1403.7458.32 ─ 14
●
select "Internal Baseband I/Q In", if the internally generated baseband signal is
used.
The RF bandwidth is reduced (see data sheet).
Alternatively, the input signal can be selected in the "I/Q Settings" menu (see
ter 5.6, "I/Q Modulation - I/Q Mod Block"
5.6.2 I/Q Settings
To open the menu for setting the I/Q modulation parameters, select "I/Q Mod > Config-
ure > I/Q Settings" or use the MENU key.
System error correction of the I/Q modulator permits precise and repeatable measure-
ments. The correction routine should be called in the case of temperature fluctuations
of several degrees. To call the routine, press the SETUP key and select "System >
Internal Adjustment " and execute the desired adjustment procedure.
It is not possible to activate an I/Q modulation with an external analog signal ("Analog
Wideband I/Q In") simultaneously with:
●
AM,
●
digital modulation,
●
digital standards,
●
noise,
●
arbitrary waveform generation and
●
multi carrier CW
I/Q modulation is activated and the source entered in the top section of the menu.
I/Q impairment for specific impairment of the I/Q modulation is set in the "Impairments"
section.
I/Q control can be swapped in the lower section of the menu and the optimized settings
for wideband modulation signals can be activated.
I/Q Modulation - I/Q Mod Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...