Instrument Function
R&S
®
SMJ100A
225
Operating Manual 1403.7458.32 ─ 14
Multi level lists which are filled automatically using the Fill with sensor function consist
of nine level columns, one for each attenuator stage (see
Correction List with Power Sensor Measurement Data"
Note:
Save list only after filling both columns (frequency and level), otherwise the
entries are lost.
"Frequency /
Hz"
Enters the frequency to which the level correction value applies.
Note:
The "Fill..." function allows to automatically enter any number
of frequencies with freely selectable range and increment.
Using the "Fill With Sensor" function of the "Edit" sub menu requires
only the entry of the frequency values. The level values are automati-
cally acquired by the connected power sensor.
"Power /
Power[1..9] @
HP.. 35 dB"
Enters the level correction value to which the specified frequency
applies. The values can be entered manually or automatically with the
"Fill With Sensor" function (available in the "Edit" sub menu).
"Goto"
Selects row for editing.
If Goto row is selected, a window opens for entering the requested
row.
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...