Remote Control Commands
R&S
®
SMJ100A
712
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Source>
INTernal | EXTernal | INT,EXT
INT
Uses the internally generated signal for modulation. To configure
the frequency, use the commands of the
EXT
Uses an externally applied modulation signal.
The external signal is input at the EXT MOD connector.
INT,EXT
Uses both, the internal and external modulation signals.
*RST:
INT
Example:
FM:SOUR INT
selects the internal modulation source.
Manual operation:
See
[:SOURce<hw>]:FM:STATe
<State>
Activates frequency modulation.
Note:
Activation of FM deactivates phase modulation (PM).
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
FM:STAT ON
Activates FM modulation.
Manual operation:
See
7.16.6 SOURce:FREQuency Subsystem
This subsystem contains the commands used to define the frequency settings for the
RF sources and sweeps.
[:SOURce<hw>]:FREQuency:CENTer
..............................................................................713
[:SOURce<hw>]:FREQuency:LOSCillator:MODE
.............................................................. 714
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe
................................................. 714
[:SOURce<hw>]:FREQuency[:CW|FIXed]
.........................................................................714
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL
................................................................. 715
[:SOURce<hw>]:FREQuency:MANual
..............................................................................715
................................................................................716
[:SOURce<hw>]:FREQuency:MULTiplier
..........................................................................717
[:SOURce<hw>]:FREQuency:OFFSet
.............................................................................. 717
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:STATe
................................................ 717
.................................................................................718
................................................................................718
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...