Instrument Function
R&S
®
SMJ100A
315
Operating Manual 1403.7458.32 ─ 14
Table 5-2: PRBS generators
PRBS generator
Length in bits
Feedback to
GUI selection
9-bit
2
9
-1 = 511
Registers 4, 0
PRBS 9
11-bit
2
11
-1 = 2047
Registers 2, 0
PRBS 11
16-bit
2
16
-1 = 65535
Registers 5, 3, 2, 0
PRBS 16
20-bit
2
20
-1 = 1048575
Registers 3, 0
PRBS 20
21-bit
2
21
-1 = 2097151
Registers 2, 0
PRBS 21
The "Data Source" selection for PRBS data from the dilaog is "PN11, ...".
Simple data patterns such as binary 0 strings or 1 strings can also be used as internal
modulation data. The "Data Source" selection from the menus is "ALL 0", "ALL1". A
variable bit string with a maximum length of 64 bits can be entered in an input field by
selecting "Pattern."
5.8.1.2
Internal Modulation Data from Lists
Internal modulation data can be generated and stored in the form of binary lists. A sep-
arate file is created for each list and held on the instrument's hard disk.
The file name can be defined by the user.
The "Data Source" selection from the menus is "Data List". When this menu item is
chosen the "Select Data List" button appears; this opens the file-selection window for
data lists. A list can be created and edited in the "Data List Editor", which is accessed
from the menu via the "List Management" or "Data List Management" button.
The maximum length of a data list is determined by the size of the data list memory
(see data sheet). There is no restriction on the number of lists that can be stored.
5.8.1.3
External Synchronous Modulation Data
External Serial Modulation Data via the DATA Connector
For "Custom Digital Modulation", serial modulation data can be supplied bit-by-bit via
the DATA connector. The "Data Source" selection from the menu is "Extern Serial".
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...