Remote Control Commands
R&S
®
SMJ100A
740
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Start>
float
Range:
full frequency range
Increment: see the data sheet: Resolution of frequency setting
*RST:
1 KHz
Example:
RST*
activates all presettings.
LFO:SWE:MODE AUTO
TRIG0:SOUR SING
LFO:FREQ:STAR 1 kHz
LFO:FREQ:STOP 10 kHz
LFO:FREQ:MODE SWE
LFO:SWE:EXEC
the instrument generates a single sweep cycle from 1 kHz to 10
kHz automatically after a manual trigger event occurs
(
:LFOutput:SWEep:EXECute
or
*TRG
). The step width is 1
kHz linear, with 15 ms dell time until the signal switches to the
subsequent step.
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:STOP
<Stop>
Sets the stop frequency for the LF sweep.
Parameters:
<Stop>
float
Range:
full frequency range
Increment: see the data sheet: resolution of frequency setting
*RST:
100 KHz
Example:
LFO:FREQ:STOP 10 kHz
sets the stop frequency for the LF sweep to 10 kHz.
Manual operation:
See
[:SOURce]:LFOutput[:STATe]
<State>
Activates/deactivates the LF output.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
LFO ON
activates the LF output. The settings under
LFO:FREQ
and
LFO:SWE
become effective.
Manual operation:
See
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...