Instrument Function
R&S
®
SMJ100A
416
Operating Manual 1403.7458.32 ─ 14
"Shortest File Wins"
The resulting signal period is defined by the shortest I/Q file in the
carrier table. Only the first part of longer I/Q files is used.
"User"
The signal period can be set by the user in the "Signal Period" field.
Shorter I/Q files are repeated periodically, and only the first part of
longer I/Q files is used.
"Least Common Multiple"
The output file duration is the least common multiple of all input file
durations.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:TIME:MODE
Signal Period
Sets the signal period in Signal Duration Mode "User". Shorter I/Q files are repeated
periodically, and only the first part of longer I/Q files is used.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:TIME
Power Reference
Defines the way the individual carriers in a composed multi carrier signal are leveled.
The difference between the provided modes is especially important if signals with dif-
ferent crest factors are composed together into a multi carrier signal.
"RMS"
The individual carriers are leveled based on their RMS power and the
configured "Carrier Gain".
Example:
A multi carrier signal is composed from two waveform files.
First carrier "Gain" = 0 dB
Second carrier "Gain" = -3 dB
In the resulting multi carrier signal, the
RMS power
of the second car-
rier signal will be 3 dB lower than the RMS power of the first carrier
signal.
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...