Remote Control Commands
R&S
®
SMJ100A
734
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Level>
float
Range:
-90 to 0
Increment: 0.01
*RST:
0
Default unit: dBFs
Example:
IQ:OUTP:DIG:POW:LEV -10
sets the output signal level to -10 dBFS.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:PEP
<Pep>
Sets the peak envelope power of the digital output signal (PEP).
Parameters:
<Pep>
float
Range:
-90 to 0
Increment: 0.01
*RST:
0
Default unit: dBFS
Example:
IQ:OUTP:DIG:POW:PEP -10
sets the signal output level to -10 dBFS.
Manual operation:
See
"Peak Level - Digital I/Q Out"
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:STEP[:INCRement]
<Increment>
Sets the step width for the frequency setting if the frequency values
UP/DOWN
are used
and variation mode
SOUR:IQ:OUTP:DIG:POW:STEP:MODE USER
is selected. The
command is linked to "Variation Step" for manual operation, i.e. the command also
sets the step width of the rotary knob for "Variation Active ON".
Parameters:
<Increment>
float
Range:
0 to 100
Increment: 0.01
*RST:
1
Default unit: dB
Example:
SOUR:IQ:OUTP:ANAL:POW:STEP:INCR 50 mV
sets the step width for the level setting to 50 mV.
Manual operation:
See
"Variation Step - Digital I/Q Out"
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...