Instrument Function
R&S
®
SMJ100A
434
Operating Manual 1403.7458.32 ─ 14
"On"
Corresponds to the default state of this parameter.
The signal calculation starts simultaneously with the external trigger
event but because of the instrument’s processing time the first sam-
ples are cut off and no signal is outputted. After elapsing of the inter-
nal processing time, the output signal is synchronous to the trigger
event.
"Off"
The signal output begins after elapsing of the processing time and
starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequences
with signal duration comparable with the processing time of the
instrument.
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:EXTernal:SYNChronize:OUTPut
Trigger Delay
Sets the trigger signal delay in samples on external triggering.
This enables the R&S
SMJ to be synchronized with the device under test or other
external devices.
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:DELay
Trigger Inhibit
Sets the duration for inhibiting a new trigger event subsequent to triggering. The input
is to be expressed in samples.
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...