Remote Control Basics
R&S
®
SMJ100A
444
Operating Manual 1403.7458.32 ─ 14
Socket connections are established on a specially defined port. The socket address is
a combination of the IP address or the host name of the instrument and the number of
the port configured for remote-control. All R&S
SMJ use port number 5025 for this pur-
pose. The port is configured for communication on a command-to-command basis and
for remote control from a program.
6.1.3.4
LAN Interface Messages
In the LAN connection, the interface messages are called low–level control messages.
These messages can be used to emulate interface messages of the GPIB bus.
Command
Long term
Effect on the instrument
&ABO
Abort
Aborts processing of the commands just received.
&DCL
Device Clear
Aborts processing of the commands just received and
sets the command processing software to a defined initial
state. Does not change the instrument setting.
>L
Go to Local
Transition to the "local" state (manual control). (The
instrument automatically returns to remote state when a
remote command is sent UNLESS
&NREN
was sent
before.)
>R
Go to Remote
Enables automatic transition from local state to remote
state by a subsequent remote command (after
&NREN
was sent).
&GET
Group Execute Trigger
Triggers a previously active instrument function (e.g. a
sweep). The effect of the command is the same as with
that of a pulse at the external trigger signal input.
&LLO
Local Lockout
Disables transition from remote control to manual control
by means of the front panel keys.
&NREN
Not Remote Enable
Disables automatic transition from local state to remote
state by a subsequent remote command. (To re-activate
automatic transition use
>R
.)
&POL
Serial Poll
Starts a serial poll.
6.1.4 GPIB Interface (IEC/IEEE Bus Interface)
To be able to control the instrument via the GPIB bus, the instrument and the controller
must be linked by a GPIB bus cable. A GPIB bus card, the card drivers and the pro-
gram libraries for the programming language used must be provided in the controller.
The controller must address the instrument with the GPIB bus address (see
ter 6.1.4.2, "GPIB Instrument Address"
Characteristics
The GPIB interface is described by the following characteristics:
●
Up to 15 instruments can be connected
●
The total cable length is restricted to a maximum of 15 m; the cable length between
two instruments should not exceed 2m.
Remote Control Interfaces and Protocols
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...