Instrument Function
R&S
®
SMJ100A
205
Operating Manual 1403.7458.32 ─ 14
● Press the MENU key and select "RF/A Mod > RF Frequency > Frequency/
Phase".
The combined "RF Frequency / Phase ..." settings dialog contains the parameters
to configure the phase settings of the RF signal.
The remote commands required to define the settings are described in
ter 7.16.13, "SOURce:PHASe Subsystem"
Delta Phase
Sets the phase of the RF signal. The current phase of the signal is used as the refer-
ence. This function allows, for example, the phase of the output signal to be synchron-
ized with the phase of a signal from a second signal generator.
Remote command:
Reset Delta Phase Display
Resets delta phase value. The set phase is adopted as the new current phase, i.e. the
delta phase value is reset to 0.
Remote command:
5.5.3.2
Phase Continuous Frequency
The phase continuous mode can be enabled in the "Phase Continuous Frequency Set-
tings" section of the "Frequency" dialog. In this mode, the output sinewave is phase
continuous, that means there is no phase discontinuity or glitch when changing the fre-
quency.
At the transition from one frequency to another an intermediate frequency (switching
transient) occurs, which causes spurs (see figure below). The transient is due to
changes of the synthesizer settings (e.g. dividers, filters, phase detectors).
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...