Instrument Function
R&S
®
SMJ100A
236
Operating Manual 1403.7458.32 ─ 14
Remote command:
Level (Avg.)
Indicates the measured level value with the selected unit.
Remote command:
Unit
Selects the unit used for result display.
The power sensor provides the measured value in Watt.
In which unit the measured value is indicated is selected here and might be Watt, dBm
or dBuV.
Remote command:
Permanent Display State
Activates the permanent indication of the power measurement result in the upper right
corner of the block diagram. The instrument shows the type of sensor, the correspond-
ing connector, the measurement source and - if set - the offset.
It is possible to switch the permanent display active for several sensors. The instru-
ment indicates the values of a maximum of two sensors simultaneously, in ascending
order of connected channels. If you activate the display of a third sensor, the instru-
ment fades out the values of the last highest channel number automatically.
Remote command:
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe
Display Priority
Determines whether the instrument displays the measured average or the peak power
permanently on the screen.
To select the peak power display, it is required that the R&S NRP-Zxx sensor supports
this feature. On power-on or connecting a sensor the average power value is set by
default.
To enable the permanent display in the block diagram, select
.
Remote command:
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority
Zero - Power Sensors
Activates the auto zero function.
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...