Parameters:
<Mode>
HDEViation | HBANdwidth
HDEViation
Provides full setting range of PhiM deviation. The range of mod-
ulation frequency is limited (see data sheet). Recommended for
low modulation frequencies and/or high PhiM deviation.
HBANdwidth
Provides maximum range of modulation bandwidth. Recommen-
ded for high modulation frequencies.
*RST:
HBANdwidth
Example:
PM:MODE LNO
selects Low Noise mode for external phase modulation.
Manual operation:
See
[:SOURce<hw>]:PM:SENSitivity?
Queries the input sensitivity of the externally applied signal for phase modulation. The
returned value reports the sensitivity in RAD/V. It is assigned to the voltage value for
full modulation of the input.
Return values:
<Sensitivity>
float
Example:
PM:DEV 1
sets a modulation deviation of 1RAD.
PM:SENS?
queries the input sensitivity at the external modulation input.
Response: 1
since the voltage value for full modulation is 1V, the resulting
sensitivity is precisely 1RAD/V.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:PM:SOURce
<Source>
Selects the modulation signal source for phase modulation.
You can use both, the internal and an external modulation signal at a time.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...