Instrument Function
R&S
®
SMJ100A
131
Operating Manual 1403.7458.32 ─ 14
In this dialog you can perform internal calibration routines.
The remote commands required to define these settings are described in
Adjust All
Performs all available internal calibration routines of the instrument.
Remote command:
:CALibration<hw>:ALL[:MEASure]?
High Accuracy Level Adjustment
Activates that the instrument performs level adjustment with high accuracy.
Remote command:
:CALibration<hw>:LEVel:HACCuracy[:STATe]
Path (Internal Adjustments)
Indicates the path for which the following adjustments are to be performed.
Remote command:
n.a.
Adjust Synthesis
Performs all adjustments which affect the frequency.
This includes adjustment of option R&S SMJ-B20, FM/PhiM Modulator.
Remote command:
:CALibration<hw>:FREQuency[:MEASure]?
Adjust Level
Performs all adjustments which affect the level. The acquired correction values
improve the settling time and the signal quality.
General Instrument Settings