Remote Control Commands
R&S
®
SMJ100A
765
Operating Manual 1403.7458.32 ─ 14
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:EMF:STAT 1
activates voltage level display.
Manual operation:
See
"Display Level as Voltage of EMF - RF Level"
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
<Amplitude>
Sets the RF level applied to the DUT.
Notes:
If specified, a level offset
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:
is included according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
In addition to numerical values, you can increase or decrease the values step by step
with the
UP
and
DOWN
according to the step width defined with
.
The RF output is activated with
OFF).
Parameters:
<Amplitude>
Minimum level ... Maximum level
Determines the RF output level.
Range:
Minimum level to Maximum level
*RST:
-30
Example:
The keywords of this command are largely optional. Therefore,
both the long and short form of the command are shown.
SOUR:POW:LEV:IMM:AMPL 15
or
:POW 15
sets the RF level at output A to 15 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
Note:
The level offset is also effective for level sweeps!
Specifies the constant level offset of a downstream attenuator/amplifier. If a level offset
is entered, the level entered with
:POWer
no longer corresponds to the RF output
level.
The following correlation applies:
POWer
= RF output level +
POWer:OFFSet
.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...