Remote Control Commands
R&S
®
SMJ100A
515
Operating Manual 1403.7458.32 ─ 14
:BLER:SETup:CORDer
<COrder>
Sets the byte order of the checksum (CRC).
Parameters:
<COrder>
LSB | MSB
LSB
The checksum starts with the least significant byte.
MSB
The checksum starts with the most significant byte.
*RST:
LSB
Example:
BLER:SET:CORD MSB
the checksum starts with the most significant byte.
Manual operation:
See
:BLER:SETup:DATA[:POLarity]
<Polarity>
:BERT:SETup:DATA[:POLarity]
<Polarity>
Sets the polarity of the feedback data bits.
Parameters:
<Polarity>
NORMal | INVerted
NORMal
High level stands for a logic 1, low level for a logic 0.
INVerted
Low level stands for a logic 1, high level for a logic 0.
*RST:
NORMal
Example:
BERT:SET:DATA INV
sets inversion of data signal.
Manual operation:
See
:BLER:SETup:DENable
<DEnable>
:BERT:SETup:DENable
<DEnable>
Activates/deactivates the use of the
Data Enable
signal and the polarity of the signal if
it is used. The
Data Enable
signal marks the data that is actually to be evaluated for
the BER measurement. Any data in addition to the PRBS sequence is masked and
thus not evaluated (e.g. sync, preambles, other channels, etc that are present in the
data bits supplied by the DUT).
BERT and BLER Subsystems
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...