Preparing for Use
R&S
®
SMJ100A
21
Operating Manual 1403.7458.32 ─ 14
See also
Chapter 2.3.5.1, "Standby and Ready state"
2.1.3 Display
The display clearly shows all main settings and signal generator states.
The display is divided into the following sections:
●
Frequency and level display with info line
–
Frequency and level settings containing offset and peak envelope power
(PEP).
–
Status messages
–
Brief error messages.
To access an window with detailed information for a message, use the INFO
key.
●
Block diagram
The block diagram shows the current configuration and the signal flow in the gen-
erator with the aid of function blocks containing an on/off switch. Selecting a func-
tion block opens a list of associated setting menus. Active menus, info windows
and graphs are displayed on top of the block diagram. The block diagram can be
displayed in the foreground anytime with the
key.
●
"Winbar" with labelled softkeys
Open menus are indicated by a labelled button in the "Winbar". The buttons deter-
mine the softkey functions for front-panel control. When a button or softkey is
pressed, the associated menu is displayed either in the foreground or minimized in
the "Winbar".
If required, the WINBAR is covered by a row of buttons to which menu-specific
functions of the active menu are assigned. By pressing the WINBAR key, the "Win-
bar" can be displayed in the foreground again. This allows the user to toggle
between the button and softkey functions.
2.1.4 Setup Keys
The keys to the right of the display set parameters, select views and control the win-
dows.
2.1.4.1
Keys for Setting Parameters
These keys provide direct access to the settings in the header of the instrument and
can be used for fast changing the state of the modulation and the RF signal.
Front Panel Tour
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...