Instrument Function
R&S
®
SMJ100A
365
Operating Manual 1403.7458.32 ─ 14
This enables the R&S
SMJ to be synchronized with the device under test or other
external devices.
The parameter
displays the delay converted in time.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:DELay
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:TDELay
Actual Trigger Delay
Displays the time (in seconds) an external trigger event or a trigger event form the
other path is delayed by.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:RDELay?
Trigger Inhibit
(available on external triggering)
Sets the duration for inhibiting a new trigger event subsequent to triggering. The input
is to be expressed in symbols.
In the "Retrigger" mode every trigger signal causes signal generation to restart. This
restart is inhibited for the specified number of samples.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit
Marker Mode
Provides the settings necessary to define the marker output signal for synchronizing
external instruments.
Marker x
Marker configuration for up to 4 marker channels. The settings are used to select the
marker mode defining the shape and periodicity of the markers. The contents of the
dialog change with the selected marker mode; the settings are self-explanatory.
All regular marker signals are described in
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:DIVider
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:OFFTime
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:ONTime
Marker x Delay
Defines the delay between the marker signal at the marker outputs relative to the sig-
nal generation start.
"Marker x"
For the corresponding marker, sets the delay as a number of sym-
bols.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...