Remote Control Commands
R&S
®
SMJ100A
787
Operating Manual 1403.7458.32 ─ 14
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the sweep spacing mode. The sweep spacing for level sweeps is always lin-
ear.
Return values:
<Mode>
LINear
*RST:
LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result: LIN
linear spacing
Usage:
Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
<Logarithmic>
Sets a logarithmically determined sweep step size for the RF level sweep. It is
expressed in decibels and you must enter the
value
and the unit
dB
with the command.
The level is increased by a logarithmically calculated fraction of the current level
according to:
step_size
n+1
= Level
n
+ step_size
n
x Level
n
Level
n+1
= Level
n
+ step_size
n+1
with Level
STARt
< level
STOP
, step_size =
SWE:POW:STEP[:LOG]
and n = number of
sweep steps
This parameter correlates with the number of steps
range as follows:
level_points = ((Level
STOP
- Level
STARt
) / step_size) + 1)
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
Parameters:
<Logarithmic>
float
Increment: 0.01
*RST:
1
Example:
SWE:POW:STEP 10dB
sets the step width for logarithmic sweep spacing to 10 dB of the
previous level in each instance (for a level sweep).
Manual operation:
See
[:SOURce<hw>]:SWEep:RESet[:ALL]
Resets all active sweeps to the starting point.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...