How to restart an interrupted update of an R&S NRP sensor
An accidental removal of the sensor during the update process interrupts the update.
If no other sensor is connected to the instrument, proceed as follows to restart the
update process:
1. Do not reconnect the sensor but keep it ready to be connected.
2. In the "Setup > System > NRP-Z Info Update" dialog, select "Rescue" in the left
sensor selection field
3. Activate "Run Update".
4. Confirm query in message box
5. Connect sensor within 4 seconds
The update starts, a bar informs about the progress.
Current Sensors
Shows the sensors that are connected to the generator with information on serial num-
ber, the revision state and some features.
Remote command:
Update
Section "Update" provides access to the file system in order to select a file for an R&S
NRP sensor update (Button "Select NRP File"), the selected file is indicated to the left
of the button. On the left side, the sensor to be updated is selected.
Button "Run Update" starts the update.
Note:
If the update is interrupted for example by accidental removal of the sensor dur-
ing the process, the button "Rescue" appears. Thus, you can restart the update proc-
ess.
Prerequisite is that no other sensor is connected to the instrument.
Refer to
"How to update an R&S NRP sensor"
interrupted update of an R&S NRP sensor"
on page 137 for detailed instructions.
Remote command:
n.a.
5.2.3.6
Update
After a firmware update it is occasionally required to also update the "PCI-FPGA". This
is enabled in the "Update" dialog.
At the first start of the new firmware, a message appears during the boot process if a
"PCI-FPGA" update is required. Execute a PCI-FPGA update by pressing the "PCI-
FPGA" button.
General Instrument Settings
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...