Instrument Function
R&S
®
SMJ100A
136
Operating Manual 1403.7458.32 ─ 14
5.2.3.5
NRP-Z Info/Update
The "NRP-Z Info/Update…" dialog covers information on connected power sensors,
like serial number, revision state and features of the particular sensor. You can directly
perform an update of the sensor firmware.
► To access this dialog, select "Setup > System > NRP-Z Info/Update…".
The "NRP-Z Info / Update" dialog indicates the connected R&S NRP-Z Power Sen-
sors with specific information and contains the functions to update the firmware of
a connected sensor.
The remote commands required to remotely configure the power sensor settings are
described in
Chapter 7.15, "SENSe, READ and INITiate Subsystems"
How to update an R&S NRP sensor
To perform an R&S NRP sensor update proceed as follows:
1. Open the R&S website
Voltmeters > R&S NRP-Z Sensors".
2. Select the respective sensor, e.g. R&S NRP-Z81.
3. Select "Downloads > Firmware" and the offered firmware suitable for your sensor.
4. Transfer and save the firmware on the instruments, for example in the
D:\user\
directory.
5. Connect the sensor to the R&S
SMJ and select "Setup > System > NRP-Z Info
Update" to open the dialog.
6. Select the sensor in the left sensor selection field.
7. Select the update file with "Select NRP File".
8. Start the update procedure with "Run Update".
The update starts and a bar indicates the progress.
General Instrument Settings
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...