Instrument Function
R&S
®
SMJ100A
193
Operating Manual 1403.7458.32 ─ 14
Example:
50 bits were set to "0" by the DUT. These 50 bits plus the preceding "0" are ignored in
the bit error rate measurement.
"Off"
Pattern Ignore is not active.
"All 1"
Bit sequences consisting of 32 or more consecutive "1" data are not
used (i.e. ignored) for the BER measurement.
"All 0"
Bit sequences consisting of 32 or more consecutive "0" data are not
used (i.e. ignored) for the BER measurement.
Remote command:
5.4.11 Block Error Rate Menu
The "Block Error Rate" menu for configuring the block error rate measurement is called
either in the "BERT" function block or via the MENU button under "BERT."
The top section of the "Block Error Rate" menu is used to switch on the block error rate
measurement and display the results and current status of the measurement.
The "Configuration" section is used to select the type of measurement and the termina-
tion criteria and to set the polarity of the input signals.
Bit and Block Error Rate Measurements - BERT Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...