"Sawtooth"
The sweep runs from the start to the stop frequency. Each subse-
quent sweep starts at the start frequency, that means the shape of
the sweep sequence resembles a sawtooth.
"Triangle"
The sweep runs from start to stop frequency and back, that means
the shape of the sweep resembles a triangle. A subsequent sweep
starts at the start frequency.
Remote command:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SHAPe
Step Lin/Log - LF Sweep
Sets the step width for the individual frequency sweep steps.
At each step this value is added to the current frequency.
Depending on the
mode you have set, the corresponding parameter is dis-
played.
"Step Lin"
The step width is a constant value in Hz.
Remote command:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]
"Step Log"
The step width is determined logarithmically in %, that means as a
constant fraction of the current frequency.
Remote command:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
Dwell Time - LF Sweep
Defines the duration of the individual sweep steps.
The "Dwell Time" set by the user is used as the step time of the sweep. The effective
net dwell time is shorter, reduced by the setting time. This setting time may be greater
than the time specified in the data sheet.
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...