Remote Control Basics
R&S
®
SMJ100A
479
Operating Manual 1403.7458.32 ─ 14
Table 6-5: Meaning of the bits used in the questionable status register
Bit No.
Meaning
0–15
Not used
6.5.6 Operation Status Register (STATus:OPERation)
This condition part contains information on the actions currently being performed by
the instrument, while the event part contains information on the actions performed by
the instrument since the last readout of the register.
To read the register, use the query commands
Table 6-6: Meaning of the bits used in the operation status register
Bit No.
Meaning
0
Calibrating
The bit is set during the calibration phase.
1–2
Not used
3
Sweeping
This bit is set during a sweep in automatic or single mode.
4–15
Not used
6.5.7 Application of the Status Reporting System
The purpose of the status reporting system is to monitor the status of one or several
devices in a measuring system. To do this and react appropriately, the controller must
receive and evaluate the information of all devices. The following standard methods
are used:
●
Service request
(SRQ) initiated by the instrument
●
Serial poll
of all devices in the bus system, initiated by the controller to find out
who sent an SRQ and why
●
Query of a
specific instrument status
by commands
●
Query of the
error queue
6.5.7.1
Service Request
Under certain circumstances, the instrument can send a service request (SRQ) to the
controller. Usually this service request initiates an interrupt at the controller, to which
the control program can react appropriately. An SRQ is always initiated if one or sev-
eral of bits 2, 4 or 5 of the status byte are set and enabled in the SRE. Each of these
bits combines the information of the error queue or the output buffer. To use the possi-
bilities of the service request effectively, all bits should be set to "1" in the enable regis-
ters SRE and ESE.
Status Reporting System
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...