Remote Control Commands
R&S
®
SMJ100A
769
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Power>
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST:
-30
Example:
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer:STARt
<Start>
Sets the start level for the RF sweep.
Note:
You can select any level within the setting range. The range is defined by this
start value and the
A defined offset (
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
affects the level values according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Start>
float
Determines the first level value of the sweep setting range.
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
-30
Example:
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output
A.
Manual operation:
See
[:SOURce<hw>]:POWer:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the level step by step with this step size, use the
POW:UP
and
POW:UP
com-
mands.
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the level accordingly, when you work in user-defined step
mode.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...