Instrument Function
R&S
®
SMJ100A
217
Operating Manual 1403.7458.32 ─ 14
The offset-free level, level offset and level limit are set in the top section of the
menu. The attenuator mode is set in the "Attenuator Settings" section.
In section "User Variation", you can determine the step size for adjusting the level
with the rotary knob (with "Variation Active On").
The power-on behavior of the instrument and the level display in the display
header are set in the "Power-On / EMF Settings" section (see
The lower sections provide access to the "Automatic Level Control" settings and to
"User Correction" settings. For more information about these functions, see
ter 5.5.7, "Automatic Level Control - ALC"
The remote commands required to define the settings are described in
ter 7.16.15, "SOURce:POWer Subsystem"
Level Settings
The offset-free level, attenuation mode, level offset and level limit are set in the top
section of the menu.
Amplitude
Sets the RF level of the RF output connector.
The level entered and displayed here corresponds to the level at the RF output, that
means any offset entry is not taken into consideration.
Note:
Status bar values blanked?
For security concerns or certain operating modes, the instrument blanks the frequency
and level display in the status bar.
●
The display has been disabled for security reasons, see
.
●
The display is disabled when list mode is running, see
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...