![R&S SMJ100A Operating Manual Download Page 576](http://html1.mh-extra.com/html/rands/smj100a/smj100a_operating-manual_3086611576.webp)
Remote Control Commands
R&S
®
SMJ100A
560
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Continuous>
0 | 1 | OFF | ON
*RST:
OFF
Example:
INIT:CONT ON
switches local state of continuous power measurement on.
Manual operation:
See
:READ<ch>[:POWer]?
The command triggers the measurement with power sensors and provides the power
measurement result of the selected power sensor. The value is provided with the unit
set with command
SENSe:UNIT[:POWer]
.
For certain power sensors, e.g. R&S NRP-Z81, two values are returned, first the value
for the average level and - separated by a comma - the peak level
Note:
The local state is not influenced by this command, measurements results can be
retrieved with local state on or off. For long measurement times it is recommended to
use a SRQ (MAV bit) for command synchronization.
Suffix:
<ch>
.
1..3
Return values:
<Power>
string
Example:
SENS:UNIT DBM
selects unit dBm for presentation of measurement result.
READ1?
queries the measurement result of the sensor connected to the
SENSOR interface.
Response: -45.6246576745440230
-45.6 dBm were measured at the given frequency.
or e.g. for R&S NRP-Z81
Response:
-55.62403263352178,-22.419472478812476
-55,6 dbm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency
Usage:
Query only
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
<State>
The command activates the use of the s-parameters correction data of the selected
power sensor.
Note:
For power sensor with attenuator this command is automatically set to ON.
SENSe, READ and INITiate Subsystems
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...