Remote Control Commands
R&S
®
SMJ100A
737
Operating Manual 1403.7458.32 ─ 14
Example:
IQ:OUTP:DIG:SRAT:SOUR USER
the internally generated clock reference signal is selected.
Manual operation:
See
"Sample Rate Source - Digital I/Q Output"
[:SOURce<hw>]:IQ:OUTPut:DIGital:STATe
<State>
Switches on and off the digital I/Q output signals.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
IQ:OUTP:DIG:STAT ON
the digital I/Q output signals of path A are activated.
Manual operation:
See
7.16.10 SOURce:LFOutput Subsystem
The
SOURce:LFOutput
subsystem contains the commands for setting the LF signal
source in CW and Sweep mode and for analog modulation.
Example
The following example shows how to set an LF sweep.
1. Set the sweep range.
LFOutput:FREQuency:STARt 4 kHz
LFOutput:FREQuency:STOP 10 kHz
2. Select linear or logarithmic sweep spacing.
LFOutput:SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
LFOutput:SWEep[:FREQuency]:STEP[:LINear] 100 Hz
LFOutput:SWEep[:FREQuency]:DWELl 20 ms
4. Determine the sweep mode.
LFOutput:SWEep:MODE AUTO
5. Determine the trigger.
TRIGger0:SOURce SINGle
6. Activate the sweep.
LFOutput:FREQuency:MODE SWEep
7. Trigger the sweep (depending on the mode).
LFOutput:SWEep:EXECute
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...