WavePro 7Zi
391
WP700Zi-OM-E-RevA
The next figure shows a histogram display of 17,999 parameter measurements divided or classified into 2000
bins. The standard deviation of the histogram sigma is 6.750 ps.
The instrument’s
parameter buffer
is very effective for determining the optimal number of bins to be used. An
optimal bin number is one where the change in parameter values is insignificant, and the histogram distribution
does not have a jagged appearance. With this buffer, a histogram can be dynamically redisplayed as the number
of bins is modified by the user. In addition, depending on the number of bins selected, the change in waveform
parameter values can be seen.
In the next figure, the histogram shown in the previous figure has been recalculated with 100 bins. Note how it
has become far less jagged, while the real peaks are more apparent. Also, the change in sigma is minimal (6.750
ps compared with 6.8 ps).
SDA Overview
Serial Data Analyzer Standard and Optional Capabilities
The Serial Data Analyzer is an instrument designed to provide comprehensive measurement capabilities for
evaluating serial digital signals. In addition to the WaveShape Analysis features in the standard WaveMaster
oscilloscope, the SDA provides eye pattern testing and comprehensive jitter analysis, including random and
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...