WavePro 7Zi
39
WP700Zi-OM-E-RevA
Dual Channel Acquisition
Combining Channels
Channels can be combined to increase sample rate, memory, or both in order to capture and view a signal in all
its detail. When you combine channels, uncombined channels like EXT BNC remain available for triggering, even
though they are not displayed.
In 2-channel operation, channels 2 and 3 are made active. Channels 1
or
2, and 3
or
4 can be used in Auto
operation for 20 GS/s. When channels are paired in the following combinations, the maximum sampling rate is
doubled and the record length is greatly increased:
Channel Combinations Receiving 40 GS/s
Channels 1 and 3
Channels 1 and 4
Channels 2 and 3
Channels 2 and 4
Note
: These sample rates apply to all WavePro Zi oscilloscopes (except the 715Zi without the WPZi 1GHz 4x20
GS option).
In short, sampling can be maximized to 40GS/s using two channel combinations except for channels 1
and
2, or
channels 3
and
4 (which always yield 20 GS/s. Use best practice by choosing channel 1 or 2 for your first input,
and channel 3 or 4 for the second.
Refer to
Acquisition Modes
in the specifications for maximum sample rates.
Combining Channels Procedure
Set up channel combinations on the interface as follows:
1. Touch
Timebase
Horizontal Setup
on the menu bar and the
Horizontal Setup
dialog is shown.
2. Under
Active Channels
, touch
4
,
2
or
Auto
. The maximum sample rate is shown alongside each button.
Hardware and Software Controls
The following Basic Control topics cover the general usage of the
hardware
buttons located on the oscilloscope's
front panel and the screen control interface elements of the
software
.
Front Panel Controls
Note
: Some front panel controls correspond with screen layout controls in specific ways. For example, the
front panel general control button corresponds with the
Hardcopy
function at
Utilities Utilities Setup
Hardcopy
.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...