Operator's Manual
WP700Zi-OM-E-RevA
312
In the second, the cursor reads the difference between the 11T
top and base of 67.2 mV.
In the third plot, the cursor reads the difference between the 3T
top and base of 47.2 mV.
The last plot shows the waveform with the parameters
calculated automatically.
1. pmoda 3T paa/11T top
2. pmoda 11T paa/11T top
3. paa 3T
4. paa 11T
5. top 11T
P1 contains the ratio of P3 to P5. P2 contains the ratio of P4 to
P5.
PNUM - PIT NUMBER
Description
Pit Number provides a measurement of the number of pits or spaces of interest or both. When pnum is selected
as a parameter measurement the total number of pits and/or spaces for the selected ‘nT’ range is displayed. In
the nT Table mode the number for each ‘nT’ value is displayed.