WavePro 7Zi
347
WP700Zi-OM-E-RevA
If CustomDSO is already in Basic mode, the following dialog will be displayed:
CustomDSO Basic Mode
The
Basic
CustomDSO mode offers eight Action buttons, each of which can call a different setup when touched.
The "Action Definition" field is used to enter a CustomDSO setup file name by means of the pop-up keyboard.
The
Test Basic Mode
dialog is used to execute the setups loaded under Action Definition.
Checking the
Execute scripts asynchronously
checkbox
allows scripts to
run in a separate process. Thus, the scope application can perform another function while the script is being
executed.
By checking the
Present CustomDSO menu at powerup and when menu
closed
checkbox
, the eight CustomDSO buttons will continue to be
available at the bottom of the screen after you close the CustomDSO dialog. Furthermore, they will appear
automatically each time the DSO is powered up.
Editing a CustomDSO Setup File
If the file does not exist, touch the
Edit
button
; a dialog
will appear for you to create the file. If the file does already exist, the
Edit
button enables you to modify it. The
Edit
button allows you to edit the file that is named in the
Setup file to recall
field, and not the file of the setup
that the instrument is currently in, unless these happen to be the same.
In the example used here, three setup files were made, called CustomA.lss, CustomB.lss and CustomC.lss.
Fragments from all three are shown below.
1160
Set
CustomDSO = XStreamDSO.CustomDSO
1161
‘ CustomDSO Setup A.lss
1162 CustomDSO.ActionScript1 =
“c:\LeCroy\XStream\CustomDSO\A.lss”
1163 CustomDSO.ActionEnable1 =
False
1164 CustomDSO.ActionScript1 =
“c:\LeCroy\XStream\CustomDSO\B.lss”
1165 CustomDSO.ActionEnable1 =
True
1166 CustomDSO.ActionScript1 =
“c:\LeCroy\XStream\CustomDSO\C.lss”
1167 CustomDSO.ActionEnable1 =
True
1168 CustomDSO.ActionScript1 =
“c:\LeCroy\XStream\CustomDSO\A.lss”
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...