Operator's Manual
WP700Zi-OM-E-RevA
454
goes true, it is recommended that Read Gate be used. If Read Gate is not present, the entire waveform will
be used unless the Analyze Region cursors are enabled.
7.
If for Read Gate you selected other than
none
, Touch inside the
Gate Polarity
field and select positive or
negative polarity.
8.
Touch inside the Compare to Reference checkbox to enable comparison of the Head Signal to the
Reference. Then touch the
Store Reference
button.
9. Touch
the
Stop on SAM
checkbox if you want to stop the acquisition when the signal falls below a user-
defined SAM value. Stop on SAMWhen this mode is enabled, the DDA is placed in what is essentially a
test mode for PRML signal quality based on SAM. When in Normal acquisition mode, the DDA-5005 will
continuously acquire and analyze head signals until it finds a PRML sample value with a SAM value below
a user-specified level. When this occurs, acquisition will be stopped, and you can directly view the
locations below the selected SAM threshold. This mode of operation is particularly useful for capturing
intermittent errors.
10. Touch inside the
SAM Threshold
data entry field and enter a value from 0 to 2 , using the pop-up numeric
keypad. See
SAM
in "PRML Channel Emulation" then touch
Back
to complete this setup.
11. Touch the
Show ML Markers
checkbox to enable the markers. ML MarkersThe ML Markers indicate the
location of the ideal PRML sample values based on the DDA-5005’s channel emulation.
12. Touch the
Show Level Markers
checkbox to enable the markers. Level MarkersThe Level Markers
indicate the vertical position of the PRML levels based on the amplitude of the acquired PRML signal. The
Level Markers indicate the vertical position of the PRML levels based on the amplitude of the acquired
PRML signal. They reflect the levels at the center of the display.
13. Touch the
Specify Region
checkbox to specify a start and end time, if desired. This may be necessary if
you are not using a Read Gate. If you are not using Read Gate, the Analytical Region must start with a
preamble for VCO synchronization. Touch inside the Start and End data entry fields and enter starting and
ending time values from -1.0 ks to +1.0 ks, using the pop-up numeric keypad.
14. To jump to a position in the head signal, touch inside the "Position"
Segment
field and enter a value from 1
to 999, using the pop-up numeric keypad. Then touch inside the
Byte
field and enter a value from 50 to
50,000.
15. To jump to a "worst SAM" area, after channel emulation, touch inside the
Worst Error #
field and enter a
value from 1 to 100, using the pop-up numeric keypad.
Channel Setup
1. Touch
the
Channel Setup
tab
. The
Channel Setup buttons and data entry fields appear.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...