Operator's Manual
WP700Zi-OM-E-RevA
148
6. Touch
inside
the
Demodulation
field and select a modulation mode.
7. Touch
inside
the
Carrier Freq.
field and enter the frequency of the carrier wave, using the pop-up keypad.
8. Touch
inside
the
Decimate by
field and enter a value by which to "thin out" your input waveform.
Decimation will speed up processing.
9.
If you selected
Phase
or
Frequency
modulation from the
Demodulation
menu, touch inside the
Scale
field and enter a value to rescale the waveform to. The units for phase modulation (PM) will be in radians
per division, and Hz per division for frequency modulation (FM).
10. Touch inside the
Bandwidth
field and enter the bandwidth of your source waveform.
11. Touch inside the
Max # taps
field and enter a value for the maximum allowable filter taps.
Note:
The actual number of taps used by the filter depends on the sampling rate and bandwidth selected. With
the sampling rate held constant, the number of filter taps decreases as the bandwidth increases. With the
bandwidth held constant, the number of filter taps increases as sampling rate increases.
Fast Wave Port Introduction
Fast Wave Port is a processing function for the LeCroy X-Stream family of digital oscilloscopes that enables you
to insert your own custom processing algorithm, written in the C/C++ language, into the DSO's processing stream.
Fast Wave Port maximizes data throughput from the acquisition system to your processing function. It also makes
it simple to create these custom processing functions.
The technology that makes this system possible is the ability of two processes in a Windows system to share a
region of memory. This enables the transfer of data at high-speed between the acquisition software and the
custom processing function, which runs in a separate process from the DSO application. A major benefit of Fast
Wave Port is that your application may be implemented and, more importantly, debugged independently of the
main application.
It is important to note that the transfer of the results of your processing function back into the X-Stream
processing stream is optional. If performance is the primary goal, and display or further processing of the results
within the DSO software is not required, then this may be skipped.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...