fwxx
Full Width at xx% Maximum
Histogram Amplitude
hist ampl
Histogram Amplitude
Definition:
The difference in value of the two most populated peaks in a histogram. This parameter is useful
for waveforms with two primary parameter values, such as TTL voltages, where hampl would
indicate the difference between the binary `1' and `0' voltage values.
Description
:
The values at the center (line dividing the population of peak in half) of the two highest peaks are
determined (see
pks
parameter description:). The value of the leftmost of the two peaks is the
histogram base (see
hbase
). While that of the rightmost is the histogram top (see
htop
). The
parameter is then calculated as: hampl = htop – hbase
Example:
In this histogram, hampl is 152 mV – 150 mV = 2 mV.