WavePro 7Zi
417
WP700Zi-OM-E-RevA
Frequency
100 kHz
to 4.5
GHz
The bit rate of the signal under test. This number can also be set to a sub-multiple
of the bit rate to show more than one eye. For example, 2.5 GHz will show one eye
of a 2.5 Gb/s signal while setting the control to 1.25 GHz will display two eyes.
Auto Scale
Vertical
Adjusts the vertical scale so that the one level is centered on 2.5 divisions above
the screen center and the zero level is set to -2.5 divisions.
Auto clear
sweeps
Clears the eye pattern display on each new acquisition.
Clear
Sweeps
Clears the current persistence map from the display.
Scale Top to
4 V to -4
V
Scales the one level to a specific voltage on the display.
Scale
Bottom to
4 V to -4
V
Scales the zero level to a specific voltage on the display.
SDA Theory
The SDA operates by processing a long signal acquisition. The processes include clock recovery, eye pattern
computation, jitter measurement, and bit error testing. All of these operations are performed on the same data
record. The processes are described in detail in this section.