WavePro 7Zi
401
WP700Zi-OM-E-RevA
5. Touch
the
Show Error Map
checkbox to enable a display of the error map. The error map displays bit
errors in a three-dimensional display, with errors shown in white and correct bits in dark gray. The display
is a 2-color surface map with frame number in the Y direction and bit number in the X direction. A frame is
indicated by the frame sync bit sequence and is composed of the bits from the end of one frame sync to
the start of the next one. If frame sync bits are not specified, the bits are laid out in fixed-length rows,
starting at the top left corner of the screen and proceeding from left to right and top to bottom in a raster
pattern.
6. Touch
the
Show Params
checkbox to display parameters BER, Nbits, False0, and False1.
7.
You can use the fields under
Frame
, to align the same bits one on top of the other so that bit failures will
appear as vertical lines.
Touch inside the
Mode
field and make a selection from the pop-up menu.
8.
If you selected
Header
or
Size and Header
, touch inside the
Frame header
field and enter a hexadecimal
number representing the prefix before the actual data. This prefix will be ignored and only the data will be
examined.
If you selected
Size
or
Size and Header
, touch inside the
Frame size
field and enter the number of bytes
in the frame, using the pop-up keypad.
Frame size
divides your waveform into equal pieces of the size that
you enter.
Jitter Setup
Jitter measurements are enabled by touching the Jitter button once the signal is set up. The Jitter dialog contains
four display buttons and four measurement buttons. The displays include
Bathtub
,
Jitter histogram
,
ISI
, and
Filtered Jitter
. You can switch display modes without re-acquiring the signal.
The Measurement buttons control which jitter measurements are displayed. The
Basic
button displays the total
jitter (Tj), Random jitter (Rj), and deterministic jitter (Dj). The
Dj Breakdown
button displays the components of Dj
--- periodic jitter (Pj), duty cycle distortion (DCD), and data dependent jitter (DDj) --- in addition to the basic Rj, Dj,
and Tj. The
Advanced
button replaces the Tj, Rj, Dj display with the peak-to-peak and rms values of the filtered
jitter.
The SDA measures jitter by determining the timing error between the edge crossings of the data signal under test
and a reference clock, which can be derived either from the data or an external clock. The measurement can also
be performed directly relative to the edges of the external clock.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...