Operator's Manual
WP700Zi-OM-E-RevA
194
Parameter
Description
Definition
Notes
result.
Width@level
(wid@lv)
Width measured at a user-
specified level.
Time between two
transitions of opposite
slope at a specified level.
(Slope specified for 1st
transition.)
Enhanced version sets
measurement calculations
to use one of the
following:
Base & Top (% or
absolute)
Peak-Peak (%)
0V-Max (%)
0V-Min (%)
Reference levels and edge-
transition polarity can be selected.
Hysteresis argument used to
discriminate levels from noise in
data.
Available with JTA2, USB2, EMC,
SDA, and XMAP options.
Standard in SDA100G and
WavePro 7000A scopes.
Enhanced parameter available with
EMC option.
WidthN
(widn)
Width measured at the 50%
level and negative slope.
Standard parameter.
X@max
Determines the horizontal
axis location of the maximum
value between the measure
gate.
Restricted to time and frequency
waveforms only.
Standard parameter.
X@min
Determines the horizontal
axis location of the minimum
value between the measure
gate.
Restricted to time and frequency
waveforms only.
Standard parameter.
Zero Level
(zero)
Zero level of an eye diagram.
Available with SDA and SDM
options.
Standard in SDA, SDA100G, and
WaveExpert scopes.
Qualified Parameters
Some LeCroy instruments and software packages give you the ability to constrain parameter measurements to a
vertically or horizontally limited range, or to occurrences gated by a second waveform. Furthermore, both
constraints can operate together. This capability enables you to exclude unwanted characteristics from your
measurements. It is much more restrictive than
Measure Gate
, which is used only to narrow the span of the
waveform for analysis, along the horizontal axis.
Note:
Since this feature operates on only a subset of the data, possible alerts or status indicators concerning
the measurement (such as " Data range too low") are not displayed.
Range Limited Parameters
Setting Up Range Qualifiers
1. Touch
Measure Measure Setup
on the menu bar.
2. Touch
a
Px
tab to open the setup dialog.
3. Touch
inside
the
Source
field and select a source from the pop-up menu.
4. Touch
inside
the
Measure
field and select a parameter from the pop-up menu.
5. Touch
the
Accept
tab of the dialog on the right, then touch the
Values In Range
checkbox.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...