![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 163](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310163.webp)
Operator's Manual
WP700Zi-OM-E-RevA
162
dialog
for
P1.
Touch the row titles
to display the top
Measure
dialog.
Note
: Some instruments have only one
Measure
dialog. In this case, touching anywhere in the parameter
results area displays the
Measure
dialog.
Parameter Setup
1.
Touch the parameter tab (
Px
) of an unused location or one that you want to change.
2. Under
Type
, select a measurement type:
Measure On Waveforms
- Applies measurements directly on the waveform denoted in the
Source1
field.
Math On Parameters
- Performs math (addition, subtraction, multiplication, division) on parameters
denoted in the
Source1
and
Source2
fields. This feature is available with some optional math
packages.
Advanced Web Edit
- Uses LeCroy's Processing Web to set up the measurement. This feature,
which is available with the XWEB option, allows you to chain practically unlimited math functions
together to operate on your waveform measurements.
3. Touch
inside
the
Source1
field and select a channel, math trace, memory trace, or other waveform to be
measured.
4.
If you selected
Measure On Waveforms
, touch inside the
Measure
field and select a parameter from the
pop-up menu
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...