Operator's Manual
WP700Zi-OM-E-RevA
528
AC Power Source
100 to 240 Vrms (+/-10%) AC at 50/60 Hz; 115 Vrms (+/-10%) AC at 400 Hz; Automatic AC voltage selection;
Installation Category: 300V CAT II
No manual voltage selection is required because the instrument automatically adapts to line voltage.
WavePro 700Zi
- </= 800 watts (800 VA) depending on accessories installed (probes, PC port plug-ins, etc.).
Note
: The instrument automatically adapts itself to the AC line input within the following ranges:
Voltage Range:
90 to 264 Vrms
90 to 132 Vrms
Frequency Range:
45 to 66 Hz
360 to 440 Hz
Power and Ground Connections
The instrument is provided with a grounded cord set containing a molded three-terminal polarized plug and a
standard IEC320 (Type C13) connector for making line voltage and safety ground connection. The AC inlet
ground terminal is connected directly to the frame of the instrument. For adequate protection against electrical
shock hazard, the power cord plug must be inserted into a mating AC outlet containing a safety ground contact.
WARNING - Electrical Shock Hazard
Any interruption of the protective conductor inside or outside of the oscilloscope, or disconnection of the safety
ground terminal creates a hazardous situation.
Intentional interruption is prohibited.
In Standby mode the oscilloscope is still connected to the AC supply. The instrument can only be placed in a
complete Power Off state by physically disconnecting the power cord from the AC supply.
The oscilloscope should be positioned to allow easy access to the socket-outlet. Disconnect the oscilloscope from
the AC supply by unplugging the instrument’s power cord from the AC outlet after the oscilloscope is placed in
Standby state.
CAUTION
The outer shells of the front panel terminals (CH1, CH2, CH3, CH4, AUX IN, AUX OUT) are connected to the
instrument’s chassis and therefore to the safety ground.
Standby (Power) Switch and Oscilloscope Operational States
The front Standby (Power) switch controls the operational state of the oscilloscope. This toggle switch is activated
by momentarily pressing and releasing it. The color of the LED below the switch indicates the status of the
oscilloscope as follows:
x
On (LED Green)*
– oscilloscope is fully powered and operational
x
Standby (LED Blinks Green)*
– oscilloscope is powered off (except for some housekeeping circuits)
x
Standby (LED Blinks Green)
– oscilloscope’s computer subsystems (hard drive, etc.) are in Standby
(reduced Power mode). All other oscilloscope subsystems are fully powered.
* Factory Settings
The oscilloscope’s factory settings result in only two basic oscilloscope states: On (LED Green) or Standby (LED
Off). In this case of Standby (LED Off), the oscilloscope is powered off with the exception of some housekeeping
circuitry (approximately 12 watts dissipation). The oscilloscope can only be placed in a complete power off state
by unplugging the instrument’s power cord from the primary power source (AC outlet). It is recommended that the
power cord be unplugged from the AC outlet if the oscilloscope is not being used for an extended period of time.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...