WavePro 7Zi
143
WP700Zi-OM-E-RevA
This is also the formula used to compute summed averaging. But by setting a "sweeps" value, you establish a
fixed weight that is assigned to the old average once the number of "sweeps" is reached. For example, for a
sweeps (weight) value of
4
:
1
st
sweep (no old average yet):
new average = (new data +0 * old average)/(0 + 1) = new data only
2
nd
sweep:
new average = (new data + 1*old average)/(1 + 1) = 1/2 new data +1/2 old average
3
rd
sweep:
new average = (new data + 2 * old average)/(2 + 1) = 1/3 new data + 2/3 old average
4
th
sweep:
new average = (new data + 3 * old average)/(3 + 1) = 1/4 new data + 3/4 old average
5
th
sweep:
new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
6
th
sweep:
new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
7
th
sweep:
new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average
In this way, for sweeps > 4 the importance of the old average begins to decrease exponentially.
Note:
The number of sweeps used to compute the average will be displayed in the bottom line of the trace
descriptor label:
Setting Up Continuous Averaging
1. Touch
Math
Math Setup
on the menu bar.
2.
Select a function tab from
F1
through
Fx
[The number of math traces available depends on the software
options loaded on your scope. See Specifications.]
3. Touch
inside
the
Source1
field and select a source waveform from the pop-up menu.
4. Touch
inside
the
Operator1
field and select
Average
from the
Select Math Operator
menu.
5. Touch
the
Average
tab in the dialog to the right of the "Fx" dialog, touch the
Continuous
button.
6. Touch
inside
the
Sweeps
data entry field and enter a value using the pop-up keypad. The valid range is 1
to 1,000,000 sweeps.
Setting Up Summed Averaging
1. Touch
Math
Math Setup
on the menu bar.
2.
Select a function tab from
F1
through
Fx
[The number of math traces available depends on the software
options loaded on your scope. See Specifications.]
3. Touch
inside
the
Source1
field and select a source waveform from the pop-up menu.
4. Touch
inside
the
Operator1
field and select
Average
from the
Select Math Operator
menu.
5. Touch
the
Average
tab in the dialog to the right of the "Fx" dialog, then touch the
Summed
button.
6. Touch
inside
the
Sweeps
data entry field and type in a value using the pop-up keypad. The valid range. is
1 to 1,000,000 sweeps.
Enhanced Resolution
ERES (Enhanced Resolution) filtering increases vertical resolution, allowing you to distinguish closely spaced
voltage levels. The functioning of the instrument's ERES is similar to smoothing the signal with a simple, moving-
average filter. However, it is more efficient concerning bandwidth and pass-band filtering. Use ERES on single-
shot waveforms, or where the data record is slowly repetitive (when you cannot use averaging). Use it to reduce
noise when your signal is noticeably noisy, but you do not need to perform noise measurements. Also use it when
you perform high-precision voltage measurements: zooming with high vertical gain, for example.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...