![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310057.webp)
Operator's Manual
WP700Zi-OM-E-RevA
56
x
Edge
- for detecting the occurrence of edges; selectable slope and level
x
Non-monotonic
- for detecting threshold re-crosses; selectable slope, hysteresis, and level
x
Runt
- for detecting pulses that fail to cross a threshold; selectable polarity and thresholds
x
Measurement
- for filtering and performing parameter measurements
Parameter Measurements
Besides parameter measurements made during acquisition, post-acquisition measurements can also be made.
The number of parameters available depends on the options loaded on your instrument. Measurements are made
only on the events defined by your filter (search criteria). A
Filter Wizard
is provided to quickly establish statistical
criteria such as ±1, 3, or 5 sigma.
Sampling Mode
Whenever WaveScan is enabled, the instrument reverts to Real-time sampling mode.
Source View
The top trace on the screen is the source (channel, math, memory) trace. This trace shows all captured edges
and highlights those that fit your search criteria. For example, in this figure we are searching for slow rising edges
that fall outside a time window:
In this acquisition, WaveScan has located eleven edges that fit the search criteria (filter) of greater than 1.2 ns
with a delta of 50 ps. Each of the found edges is highlighted with a red rectangle, and indexed to a table entry at
left. Rise time values are also included in the table.
Level Markers
Markers are provided on the source trace to show measurement criteria such as level and thresholds. In the
example above, the level markers indicate 10% and 90% for the standard Rise Time parameter measurement.
Level markers are displayed only while the scope is in WaveScan mode. Once the WaveScan
Enable
checkbox
is unchecked, the level markers disappear.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...