WavePro 7Zi
241
WP700Zi-OM-E-RevA
The next waveform figure has software-assisted triggering enabled. The red trace is aligned with the trigger point,
indicated by the red marker below the grid. The white trace is the waveform at the input to software-assisted
trigger processing. Notice how the white trace is the same as the previous figure.
The next figure shows a waveform outside the horizontal gate region (the closest positive edge is not /-1
ns). The
Auto
and
Normal
difference is as follows:
x
In
Auto
software assisted trigger mode, the waveform would be displayed, but not aligned.
x
In
Normal
mode, it would not be displayed.
Edge Trigger on Simple Signals
The instrument uses many waveform capture techniques that trigger on features and conditions that you define.
These triggers fall into two major categories:
x
Simple Triggers
– activated by basic waveform features or conditions such as a positive or negative
slope, and hold-off
x
SMART Triggers
– sophisticated triggers that enable you to use basic or complex conditions for triggering.
Use SMART Triggers for signals with rare features, like glitches.
Trigger Settings
Horizontal and Vertical adjustments are typically made for all trigger types using either the
Delay
or
Level
knobs
on the front panel of the instrument or their respective fields on the
Timebase
setup dialog.
x
Horizontal:
Turn the
D
ELAY
knob in the HORIZONTAL control group to adjust the trigger's horizontal
position. Or, touch inside the
Delay
field in the timebase setup dialog and enter a value, using the pop-up
keypad.
The trigger location is shown by a marker marker
under the grid.
Post-trigger delay is indicated by a left-pointing arrow
at the lower-left of the grid. The time value is
given in the title line of the
TimeBase
label
at the lower-right of the grid.
x
Vertical:
Turn the
L
EVEL
knob in the TRIGGER control group to adjust the trigger's vertical threshold.
Turn this knob to adjust the level of the trigger source or the highlighted trace. Level defines the source
voltage at which the trigger will generate an event: a change in the input signal that satisfies the trigger
conditions.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...