WavePro 7Zi
177
WP700Zi-OM-E-RevA
Parameter
Description
Definition
Notes
DCD
Amount of jitter due to duty
cycle distortion
Available with SDA and ENET
options.
Standard in SDA and WaveExpert
scopes.
DDj
Amount of data dependent
jitter in a signal
Available with SDA option.
Delay
Time from trigger to
transition: Measures time
between trigger and first 50%
crossing after left cursor. Can
measure propagation delay
between two signals by
triggering on one and
determining delay of other.
Time between trigger and
first 50% crossing after
left cursor
Standard parameter.
Delta Delay
(ddelay)
Computes time between 50%
level of two sources.
Time between midpoint
transition of two sources
Standard parameter.
Dj Effective
(Dje)
Amount of deterministic jitter
(estimated) in a signal
Available with SDA option.
DOV
Differential Output Voltage of
a 100Base-T signal
Available with ENET option.
Dperiod@level
(dper@lv)
Adjacent cycle deviation
(cycle-to-cycle jitter) of each
cycle in a waveform
Reference levels and edge-
transition polarity can be selected.
Hysteresis argument used to
discriminate levels from noise in
data.
Available with JTA2, SDA, and
XMAP options.
Standard in SDA100G scopes.
Droop FG
(drpfg)
1000Base-T test mode 1
droop from F to G
Available with ENET option.
Droop HJ
(drphg)
1000Base-T test mode 1
droop from H to J
Available with ENET option.
Dtime@level
(dt@lv)
Computes transition between
selected levels or sources.
Time between transition
levels of two sources, or
from trigger to transition
level of a single source
This measurement gives
the time of the source 2
edge minus the time of
the source 1 edge.
Reference levels and edge-
transition polarity can be selected.
Hysteresis argument used to
discriminate levels from noise in
data.
Standard parameter.
Dtrig Time
(dtrig)
Time from last trigger to this
trigger
Standard in WaveRunner 6000A,
WavePro 7000A, WaveMaster, and
sampling scopes.
Duration
(dur)
For single sweep waveforms,
dur is 0; for sequence
waveforms: time from first to
last segment's trigger; for
Time from first to last
acquisition: for average,
histogram or sequence
waveforms
Standard parameter.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...