WavePro 7Zi
293
WP700Zi-OM-E-RevA
Beginning Edge Shift Sigma (BESS)
Description
BESS provides a measurement of the mean, normalized standard deviatio
n
of the Beginning Edge Shift
measurements (see
BES
). When a single
n
is specified, or when you are in ‘nT Table’ Show mode, the value
calculated for the
n
th index is calculated using the following equation for standard deviation:
Beginning Edge Shift Sigma cannot be calculated for a given index
n
unless there are at least two Beginning
Edge Shift values calculated or that
n
index.
When Beginning Edge Shift is configured as a custom parameter with a range of
n
, the value calculated is the
standard deviation of the distribution that results by normalizing each independent distribution categorized by the
space (pit) nT preceding the subject pit (space). Distributions are normalized by subtracting the mean of the
distribution from all of the elements in the distribution. This results in the following equation for overall Beginning
Edge Shift Sigma resulting from the individually categorized Beginning Edge Shift Sigma values:
Note
: The value calculated by BESS will generally not be the same as the sigma of the BES measurement
displayed on the parameter line when a range of
n
is used and statistics is on. This is because the two
measurements are not the same. The BESS measurement normalizes the results for each
n
by subtracting the
mean BES from each BES in the nth distribution. This results in a superposition of mean-centered distributions,
not a superposition of 0-centered distributions contributing to BES measurements. BESS will always be less
than or equal to the standard deviation of BES measurements.
Display Options
ORM parameter calculations can be displayed, histogrammed and trended in a variety of ways.
DISPLAY TYPE
VALUE DISPLAYED
Parameter
Statistics Off
Single value of the standard deviation of the mean normalized beginning edge shift
values for pits/spaces of interest for last acquisition.
Parameter
Statistics On
Average, minimum, maximum, and sigma of the beginning edge shift sigma value
calculated per acquisition for all acquisitions since the last CLEAR SWEEPS operation.
nT Table
List of values of the standard deviation of the beginning edge shift values for each ‘nT’
spaces (pit) within the selected range preceding the subject pit (space) for the last
acquisition.
Histogram
Function
Histogram of beginning edge shift sigma values calculated for each acquisition for all
acquisitions since the last CLEAR SWEEPS operation.
Trend Function
Trend of the beginning edge shift sigma values calculated for each acquisition for all
acquisitions since the last CLEAR SWEEPS operation.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...