Parameters allow measurements of the section of waveform lying between the parameter cursors. The position of
the cursors can be set by dragging, entering an exact value in the "Standard Cursors" dialog, or by means of the
C
URSORS
front panel knobs. When you enable tracking by checking the
Track
checkbox, you can move the
parameter cursors across the waveform so that measurement results can be taken on different sections of the
waveform.
BES or EES Table
When the selected measurement is Begin Edge Shift (BES), End Edge Shift (EES), or their
sigmas (BESS or EESS), the results can be shown in a table. Select
nT Table
from the
Show
menu to display a table of the average value of BES or EES for each subject nT vs. each
“preceded” or “followed” nT:
Note:
For EES, “Following nT” is displayed instead of “Preceding nT.” The nT is the range specified, starting
from the low T. The subject T (i.e., s(n) ) will also start from Low T to High T.
Setup and Measurement Dialog
The “AORM Measurements” dialog is accessed from the menu bar’s
Analysis
menu. AORM is supplied with X-
Stream software version 4.2 and later. This highly interactive dialog allows you to set up and configure the clock
and data sources, select a measurement type, and analyze the waveform, including statistics on parametric
measurements:
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...