Operator's Manual
WP700Zi-OM-E-RevA
374
ET Delay
- The time in seconds between the center of the pulse (1/2 bit) and the edge where the trigger is set
ET Center
- The center of the mask in DIV (usually 5.0)
Base Point
- Optional (currently not used)
Offset Tol
(for "Ab Offset" type only) - The permitted tolerance to adjust the offset for a better mask fit
Gate Start
- The limit in DIV in the waveform data at which the mask test is started
Gate Stop
- The limit in DIV in the waveform data at which the mask test is stopped. This parameter and the one
above it allow you to perform mask testing on specific pulses within a longer, more complex waveform.
Mask Data
(optional) - Hyperlink to the table that contains the mask data
Mask File Name
(optional) - Used if MaskData is not specified; file name of the custom mask (*.msk file). This file
type is created by LeCroy’s MaskMaker software.
The following images show graphically how the above parameters are used for the four different types of masks
set in the Type field. They show how the various parameters are used to align and scale the waveforms to the
mask.
Absolute
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...