WavePro 7Zi
307
WP700Zi-OM-E-RevA
Example
This persistence waveform is created by setting a SMART
Trigger® to capture only 4T spaces. The 4T base computed is -
27.0 mV.
When the same measurement is taken with the parameter
cursors, it confirms that -27.0 mV is a reasonable value for the
base.
PMAX - PIT MAXIMUM
Description
Pit Maximum provides a measurement of the maximum voltage value of pits of interest. It provides a comparison
of how the maximum point in the waveform corresponds to the ptop value When pmax is configured as a custom
parameter, all maximums within the single nT or range of nT are calculated. Histogramming or trending such a
configuration would result in one value per pit in the nT range contributing a value to the histogram or trend. The
value displayed on the custom parameter display line is the average of all such maximum calculations. ‘nT Table’
mode provides an average maximum value for the pits in each
n
index.