Operator's Manual
WP700Zi-OM-E-RevA
198
Moreover, a hysteresis range may be specified to ignore any spurious transition that does not exceed the
boundaries of the hysteresis interval. In the figure, Delta c2d- (1, 2) measures the time interval separating the
rising edge of the clock (trigger) from the first negative transition of the data signal. Similarly, Delta c2d+ (1, 2)
measures the time interval between the trigger and the next transition of the data signal.
Level and Slope
For several time-based measurements, you can choose positive, negative, or both slopes to begin parameter
measurements if these parameters are included in your scope. For two-input parameters, such as Dtime@level,
you can specify the slope for each input, as well as the level and type (percent or absolute).
Print, Plot, or Copy
The instrument gives you the ability to output files to a printer or plotter, to print to file, or to e-mail your files. Any
Microsoft Windows® supported printer is supported by your instrument.
Printing
Setting Up the Printer
1. Touch
File
Print Setup
on the menu bar. The Utilities
Hardcopy
dialog opens.
OR
Press the P
RINT
front panel button. Then, touch the
Print Setup
button on the
flyout menu.
2.
On the Hardcopy dialog, touch the
Printer
icon
.
3. Under
Colors
, touch the
Use Print Colors
checkbox if you want the traces printed on a white background.
Note
: A white background saves printer toner. (You can change the printer colors in the
Utilities Preference
Setup Color
dialog.)
4. Touch
inside
the
Select Printer
field. Choose a printer from the pop-up menu.
5. Touch
the
Properties
button to see your printer setup.
4.
Touch the icon for the layout
Orientation
you want: portrait or landscape.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...