Parameter
Description
Definition
Notes
Peaks
Number of peaks in a
histogram.
Available with DDM2, JTA2,
XMATH, XWAV, CAN02, SDA, and
XMAP options.
Standard in DDA-5005A, SDA100G,
and WaveExpert scopes.
Peak to Peak
(pkpk)
Peak-to-peak: Difference
between highest and lowest
points in waveform. Unlike
ampl, does not assume the
waveform has two levels.
maximum - minimum
Gives a similar result when applied
to time domain waveform or
histogram of data of the same
waveform. But with histograms,
result may include contributions
from more than one acquisition.
Standard parameter.
Percentile
(pctl)
Horizontal data value that
divides a histogram so the
population to the left is xx%
of the total.
Available with DDM2, JTA2,
XMATH, XWAV, CAN02, SDA, and
XMAP options.
Standard in DDA-5005A, SDA100G,
WaveExpert, and sampling scopes.
Per DCD
(pdcd)
Amount of jitter due to duty
cycle distortion measured on
a persistence trace.
Standard in SDA and WaveExpert
scopes.
Per Duty Cyc
(pduty)
Duty cycle measured on a
persistence trace.
Standard in SDA and WaveExpert
scopes.
Period
Period of a cyclic signal
measured as time between
every other pair of 50%
crossings. Starting with first
transition after left cursor,
period is measured for each
transition pair, with values
averaged to give final result.
Where: Mr is the number
of leading edges found,
Mf the number of trailing
edges found,
the time
when rising edge i
crosses the x% level,
and
the time when
falling edge i crosses the
x% level.
Standard parameter.
Period@level
(per@lv)
Period at a specified level
and slope for every cycle in
waveform.
Reference levels and edge-
transition polarity can be selected.
Hysteresis argument used to
discriminate levels from noise in
data.
Available with JTA2, AORM, ENET,
SDA, and XMAP options.
Standard in DDA-5005A, SDA100G,
and WavePro 7000A scopes.
Per Pulse Sym
Symmetry of RZ pulse
around eye aperture center.
Standard in SDA and WaveExpert
scopes.
Persist Area
(parea)
Area under mean
persistence trace.
Standard in SDA and WaveExpert
scopes.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...