![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 281](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310281.webp)
Operator's Manual
WP700Zi-OM-E-RevA
280
View Menu Selections
View Displays
Additional
Functions
Parameter
The source traces will be displayed along
with the custom parameters (see
Measurements Table
). If two traces are to be
displayed, dual grids will be drawn.
Statistics
: toggles the parameter statistics
on/off.
Histogram
When
Histogram
is selected, a histogram of
the selected parameter is displayed in a
second grid, and a new tab, “AORM
Histogram,” appears.
Find Center And Width
: determines the best
scaling for the histogram based on up to the last
20,000 samples collected. This occurs
automatically if the
Enable Auto Find
checkbox
is checked.
Trend
When
Trend
is selected, a trend plot of the
selected parameter is displayed in a second
grid, and a new tab, “AORM Trend,” appears.
Find Scale
: determines the best scaling for the
trend (center and height). This occurs
automatically if the
Auto Find Scale
checkbox
is checked.
XY Plot
Plots the trend of the selected measurement
vs. either the trend t@pit or pwid, as
appropriate; not available for all
measurements (see Measurement Table,
next, for details).
When
XY Plot
is selected, a trend plot of the
selected parameter is displayed in a second
grid, and a new tab, “AORM Trend Y,”
appears.
Find Scale
: determines the best scaling for
trends (center and height). This occurs
automatically if the
Auto Find Scale
checkbox
is checked.
Equalizer and PLL Dialog
Field Description
Track Clock
Check this checkbox to enable tracking.
Filter Cutoff and
Boost
An equalizer filter is applied prior to the measurements. You can adjust the cutoff
frequency and boost of the filter.
Slicer Bandwidth
The data passes through a slicer to level the data (removes the threshold due to low
frequency effects). You can set the bandwidth of the slicer.
PLL Bandwidth
If you checked the
Clock From Data
checkbox, a PLL is used to recover the clock from
the data waveform. In that case, the bandwidth of the PLL can be adjusted.
PLL Startup
Period From
Auto
When
Auto
is selected, the startup frequency will be determined by analyzing the
peaks of the input signal.
User
When
User
is selected, you can set the startup frequency yourself. This is useful
when the startup frequency cannot be reliably determined from the input signal.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...